Language:

Model MS-8100 Hematology Anlayzer

MS-8100 Hematology Analyzer
 
Technical Specification
Parameters
WBC, Lymph, Mid, Gran, Lymph%, Mid%, Gran%, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-SD, RDW-CV, PLT, MPV, PDW, PCT, WBC Histogram, RBC Histogram, PLT Histogram, P-LCR.


Principles
・ Electrical resistance for counting and SFT method for hemoglobin .

Specification
・23 parameters, with 3 part WBC differntial
・Aperture Diameter 70μm
・10 uL sampler volume
・Memory for 30,000 results including histograms
・Throughput : 35 samples/hour
・Automatic and manual calibration
・Automatic dilution when power on/off
・High voltage cleaning after each measurement
・QC method: X-B , L-J , SD, CV


Performance

Parameter

Linearity range

Precision

WBC

0.5~99.9×10 9 /L

CV < 2.0%

RBC

0.2~ 9.99 × 10 12/L

CV < 1.5%

HGB

1.0~299.9 g/L

   CV < 1.5%   

MCV

40~200 fL

  CV < 0.5%  

PLT

0~999×10 9/L

CV < 4.0%

Carryover
・WBC< 0.5% , RBC<0.5% , HGB<0.5% , PLT < 2%

Working mode
・Single counting pathway+ Independent HGB detection system


Operating Environment
 ・ Temperature 15℃~ 30℃
 ・ Humidity  10% ~ 85%


Printout
・Thermal recorder, printer
・Optional external printer

Power Requirement
175-265V ~ 50/60HZ

Dimension
390×460×400 mm (W*D*H)
Net Weight
20 Kg

 Optional
1 Printer
2. Bar code scanner

Consumption
1 Lyse
2 Diluent
3 Rinse